The automated regression suite builds a white, flat synthetic test plane at 100 pixels per centimeter. It places a 2.0 cm × 1.0 cm blue reference marker and a 3.0 cm × 1.5 cm red target on that plane, then projects the plane through three known homographies that simulate mild, rolled, and steep camera tilts. The generated image can also be written to a temporary PNG when a failed test needs visual inspection.
Run the suite with:
pytestThe optional hardware payload lives in
firmware/medmeasure_payload.ino. Install
the ArduinoJson library, upload the sketch to an Arduino-compatible board, and
connect its simulated photodiode to A0 and PWM lighting to pin 9. The
sketch streams JSON telemetry at 115200 baud and accepts commands such as
SET_BRIGHTNESS:128 over serial.
serial_bridge.py provides the Python HardwareBridge interface. It detects
Arduino, CH340, and FTDI serial adapters, returns parsed telemetry when a board
is connected, and returns mock telemetry when hardware is unavailable. The
Streamlit sidebar displays the connection state and live light telemetry, and
can send the LED brightness command to the board. If automatic detection misses
your adapter, start the app with MEDMEASURE_SERIAL_PORT set to its serial
device (for example, /dev/ttyACM0 or COM3).
The simulated red/pink target is segmented using a combined HSV and CIE Lab colour mask rather than a single red hue range. It accepts red, pink, and shadowed-red pixels while requiring Lab red chroma plus brightness-normalized red dominance. It additionally compares both measures against a resolution-scaled local neighbourhood: subtle red/pink tissue is included when it contrasts with nearby skin, while a broad, uniformly warm skin region is not treated as a target. Strongly red pixels are retained inside a large target even when its centre has little local contrast. Small resolution-scaled morphological cleanup removes noise and lighting gaps without smoothing away the target's irregular outer boundary.
The Streamlit result includes a black-and-white segmentation-mask preview: white pixels are included in the measurement. This is an inspection aid for the hackathon simulation only, not clinical validation or diagnosis.
For every simulated tilt, the tests require reference_corners to locate the
blue marker within 4 pixels of its projected corners and require
correct_perspective to restore the marker to within 3 pixels of its expected
200 px × 100 px size. analyze must recover the target's 4.50 cm² area,
3.00 cm length, 1.50 cm width, and 9.00 cm perimeter. The acceptable relative
errors are 9% for area, 4% for length, and 6% for width and perimeter; these
allow for normal rasterization and contour-boundary effects while detecting
meaningful calibration regressions.