Small beam optimizer for the section from the ion source to Cup 1 of the FLAVIA AMS beamline at CologneAMS. SIRIUS covers the whole beamline from Cup 1 to Cup 6.
python -m venv .venv
.venv/bin/pip install -e ".[dev]"
.venv/bin/mcsirius 27 --dry-run # optimize for mass 27 u against simulated hardware
.venv/bin/pytest # tests run against the simulated machineLive modes need a running FLAVIA setup and talk to the hardware over MQTT:
| Option | What it does |
|---|---|
--dry-run |
run the optimizer against simulated hardware (currently the default) |
--live-check |
connect and print readbacks, without changing source, magnet or cup selection |
--live-prepare |
move the real machine to the deterministic source and magnet starting point, without optimizing |
--live-peak-map |
scan the real analyzing magnet around the predicted mass, without changing source voltages |
mcsirius controls only the components required to maximize beam current at Cup 1:
- sputter voltage
- extraction voltage
- analyzing magnet current
- Einzel lens voltage
The user-facing input will ultimately be the ion mass. The optimizer then prepares the source, calculates the expected magnet setting, finds the beam on Cup 1 and optimizes the source and focusing settings.
- Sputter voltage: 4 to 9 kV
- Extraction voltage: 14 to 25 kV
- Einzel lens voltage: 14 to 25 kV
- Einzel lens must remain within 2 kV of extraction voltage
- Initial operating point: 4 kV sputter, 14 kV extraction, 14 kV Einzel lens
The optimizer will use a deterministic optimization chain rather than a multi-dimensional grid scan.
Hardware access, measurement and optimization logic will remain separate so the optimizer can be tested against simulated hardware before controlling FLAVIA.
The initial analyzing-magnet setpoint is calculated from ion mass and total source energy before any measured beam search is performed.
The model uses the same geometry and current/field calibration as FLAVIA:
- bending radius: 0.5 m
- total ion energy: sputter voltage + extraction voltage
- current calibration: I = (B_kG + 0.0348763) / 0.10886
This value is only the predicted center for the later local magnet search. The optimizer will not treat magnet current as an independent global parameter.
The first executable optimization chain is intentionally simple:
- calculate the expected magnet current from ion mass and source energy
- run a bounded coarse-to-fine magnet scan around that prediction
- restore the best measured magnet setting
- scan the Einzel lens while keeping it within 2 kV of extraction
- restore the best magnet and Einzel settings
- use the magnitude of the Cup 1 current as the optimization score
The scan distances and step sizes remain explicit inputs until suitable machine values have been verified experimentally.
A complete optimization pass follows a fixed, inspectable sequence:
- calculate and locally optimize the analyzing magnet
- focus the Einzel lens on Cup 1
- optimize extraction voltage
- recalculate and locally optimize the magnet for each extraction candidate
- optimize sputter voltage
- recalculate and locally optimize the magnet for each sputter candidate
- finish with another magnet and Einzel optimization
Extraction and Einzel are moved together in bounded intermediate steps so that their voltage difference never exceeds the configured machine limit.
Real optimization should not rely on one instantaneous Keithley value.
The measurement layer can therefore wrap the hardware and:
- wait for the beam to settle after a changed setting
- acquire several current samples
- use the median current as the optimization value
- calculate the median absolute deviation (MAD) as a simple noise estimate
- retain measurement history for later diagnostics
The timing and sample count remain configurable until the real Keithley update rate and useful settling times have been measured on FLAVIA.
The optimizer can repeat complete source-to-Cup-1 cycles.
A run stops when either:
- the configured maximum number of cycles has been reached, or
- the relative Cup-1 improvement from one complete cycle to the next falls below the configured threshold
The default limit is three complete cycles with a one-percent relative improvement threshold. These values are configuration defaults rather than hard machine limits.